DTEL-15G  Multifunction Process Calibrator is a high-performance instrument specially designed for calibrating thermal secondary instruments, pressure instruments and DCS systems. It is a beautiful and durable instrument, which is an ideal tool for calibrating, maintaining and overhauling thermal instruments.


Function

1. Outputs V, mV, mA, Ω, Hz, and can also simulate thermocouple or RTD signals.

2. Measures V, mV, mA, Ω, Hz, pressure, calibrates pressure switches, and can also measure temperature using thermocouples or RTDs.

3. Measurement and output can be performed simultaneously and isolated from each other.

4. Provides 24V DC power supply, isolated from measurement and outputs.

5. High-resolution display, good human-computer interaction interface;

6. Support step output, can be programmed with common step signals and step intervals, and support two modes of automatic step or manual step;
7. Using Pt100 temperature probe for cold end temperature compensation, support ambient temperature tracking or constant temperature setting two modes;
8. Multi-functional, programmable unit conversion function, can be V, mA output or measurement signals into other unit quantities (℃, ℉, % , kPa, MPa, psi, mmHg, mm H2 O, m3 / s, m3 / min, m3 / h, L / s, L  / h, L / min, m, ?) ;

9. When outputting mA signal, it supports two modes of mA source or analog transmitter;

10. When outputting and measuring thermocouple ℃ signal, display the corresponding mV value;

11. When outputting and measuring RTD ℃ signal, display the corresponding Ω value;

12. When outputting and measuring Ω signal, it supports two, three and four-wire modes;

13. When outputting Hz signal, the amplitude can be set, supporting square wave or sine wave mode;

14. Provide 24V DC circuit power supply and measure the mA signal in the circuit;

15. Can be in any digit of the output value, signal fine-tuning;

16. Can lock or unlock the measurement value when measuring the value;

17. Can collect the minimum and maximum value and calculate the average value when measuring the value;Can store on-site inspection data, use the communication software data transcription function for management;

18. Can store on-site inspection data, use the communication software data transcription function for management;

19. Strong lithium 7.4V/4800mAh, can continue to work more than 12 ~ 30 hours;

20. Output and measurement items can be calibrated.


Technical Parameters


Output Range

0.01 level Maximum 

                 allowable tolerance

0.02 level Maximum 

                  allowable tolerance

Resolution

10.99999 V

± (0.008% of reading) +0.0001 V)

± (0.015 % of reading) +0.0001 V)

10 μV

1.099999 V

± (0.008% of reading) +0.00001 V)

± (0.015 % of reading) +0.00001 V)

1 μV

– 99.9999 mV~109.9999 mV

± (0.008% of reading) +0.003 mV)

± (0.015 % of reading) +0.003 mV)

1 μV

30.0999 mA

± (0.01 % of reading) + 1 μA)

± (0.015 % of reading) + 1 μA)

0. 1 μA

4000.00 Ω

±( 0.008% reading +0.04 Ω)

± (0.015 % of reading) +0.04 Ω)

10 mΩ

400.000 Ω

± (0.008% of reading) +0.005 Ω)

± (0.015 % of reading) +0.005 Ω)

1 mΩ

6 types Thermal resistance ℃

See " RTD Thermal resistance 

             Technical specifications"

See " RTD Thermal resistance 
     Technical specifications"

0.01 ℃

54999.9 Hz

± 2 Hz

± 2 Hz

0. 1 Hz

5499.99 Hz

± 0.2 Hz

± 0.2 Hz

0.01 Hz

549.999 Hz

± 0.02 Hz

± 0.02 Hz

0.001 Hz

24V DC circuit power

± 5 %

± 5 %

 

Sine wave peak-to-peak: (0.2  ~  19.8) Vp-p Magnitude accuracy: 2 % of reading + 0. 1 % of range
Sine wave peak-to-peak: (0.2  ~  19.8) Vp-p Magnitude accuracy: 2 % of reading + 0. 2 % of range
Sine wave peak-to-peak: (0.2  ~  19.8) Vp-p Magnitude accuracy: 2 % of reading + 0. 3 % of range
Sine wave peak-to-peak: (0.2  ~  19.8) Vp-p Magnitude accuracy: 2 % of reading + 0. 4 % of range
Sine wave peak-to-peak: (0.2  ~  19.8) Vp-p Magnitude accuracy: 2 % of reading + 0. 5 % of range



TC Technical specifications (one year, ambient temperature: 20 ℃ ±5 ℃)

Thermocouple

   index number

temperature 

                   range

Maximum permissible error (analogue output and measurement)

0.01class

0.02class

E

(–100 ~ 700) ℃

± 0.09℃

± 0.14℃

(700 ~ 1000) ℃

± 0.12℃

± 0.19℃

 

J

(– 210 ~ –100)℃

± 0.19℃

± 0.22℃

(–100 ~ 700) ℃

± 0.10℃

± 0.14℃

(700  ~ 1200) ℃

± 0.15℃

± 0.23℃

T

(– 250 ~ –100 ℃

± 0.55℃

± 0.62℃

(–100  ~ 0) ℃

± 0.12℃

± 0.12℃

(0 ~ 400) ℃

± 0.08℃

± 0.10℃

WRe3-25

(0 ~ 2000) ℃

± 0.42℃

± 0.59℃

(2000  ~ 2315)℃

± 0.68℃

± 0.99℃

WRe5-26

(0 ~ 1000)℃

± 0.24℃

± 0.31℃

(1000 ~ 2000)℃

± 0.47℃

± 0.66℃

(2000 ~ 2315)℃

± 0.66℃

± 0.95℃

S

(–20 ~ 0) ℃

± 0.60℃

± 0.60℃

(0 ~ 100) ℃

± 0.56℃

± 0.56℃

(100 ~ 1768) ℃

± 0.41℃

± 0.53℃

R

(–20 ~ 0) ℃

± 0.60℃

± 0.76℃

(0 ~ 200) ℃

± 0.57℃

± 0.76℃

(200 ~ 1768) ℃

± 0.39℃

± 0.51℃

B

(600 ~ 800) ℃

± 0.56℃

± 0.58℃

(800 ~ 1820) ℃

± 0.43℃

± 0.46℃

K

(–250 ~ –200) ℃

± 0.72℃

± 0.82℃

(–200 ~ –100) ℃

± 0.23℃

± 0.25℃

(–100 ~ 600) ℃

± 0.12℃

± 0.16℃

(600 ~ 1372) ℃

± 0.22℃

± 0.33℃

N

(–200 ~ –100) ℃

± 0.33℃

± 0.36℃

(–100  ~ 1300) ℃

± 0.19℃

± 0.28℃

E

(–250 ~ –200) ℃

± 0.39℃

± 0.46℃

(–200 ~ –100) ℃

± 0.15℃

± 0.17℃



RTD index

   number

temperature 

        range

Maximum permissible error (analogue output)

Maximum permissible error (measurement)

0.01 class

0.02 class

0.01 class

0.02 class

 

Pt100

(200 ~ 200) ℃

±0.05℃

±0.09℃

±0.05℃

±0.09℃

(-20 ~ 600) ℃

±0.09℃

±0.16℃

±0.09℃

±0.16℃

(600 ~ 850) ℃

±0.13℃

±0.22℃

±0.13℃

±0.22℃

 

Pt1000

(-200 ~ 200) ℃

±0.05℃

±0.08℃

±0.05℃

±0.09℃

(200 ~ 600) ℃

±0.09℃

±0.16℃

±0.09℃

±0.16℃

(600 ~ 850) ℃

±0.12℃

±0.22℃

±0.12℃

±0.22℃

Cu50

(-50 ~ 150) ℃

±0.05℃

±0.08℃

±0.05℃

±0.08℃

 

BA1

(–200 ~ 0) ℃

±0.05℃

±0.07℃

±0.05℃

±0.07℃

( 0 ~ 400) ℃

±0.09℃

±0.14℃

±0.09℃

±0.14℃

(400 ~ 650) ℃

±0.12℃

±0.19℃

±0.12℃

±0.19℃

BA3

(–50 ~ 100) ℃

±0.06℃

±0.09℃

±0.03℃

±0.06℃

G

( –50 ~ 150) ℃

±0.05℃

±0.08℃

±0.05℃

±0.08℃


Rrandom attachment

Manual, certificate, warranty card, factory test report, rechargeable battery, charger, cold junction temperature compensation probe, 24V DC output line, test line kit, instrument box, etc., calibration certificate (optional, cost is extra)

DTEL-15G Multifunction Process Calibrator
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
This article is read 239 Times
Previous:China Nuclear E&C Groupnext:DTW-670G High Temperature Smart Calibration Salt Bath

Contact us

DEARTO QR codeHeadquarter Address Tai'an High-tech Development Zone,  Shandong Province 
Telephon :+86 0538-5089056  +86 13953856217 
Fax : +86 0538-5059718 
After sale:+86 0538-5050959